This is the third part of a three part series about portable gas detector safety. The first blog discussed the basics of portable gas detectors while the second blog discussed the how and when of testing and calibration. Thanks to a great article in OH&S, today we’ll talk about common myths about gas detector safety.
Myth 1: Gas detectors needs to be calibrated only to make up for sensor drift
Not true. Basically, any time a gas detector is used, all prior bump tests are no longer valid. Anything could have happened to the unit since it was tested and you need to check it again with gas.
Myth 2: My gas detector does a self bump test so I don’t need to calibrate or test it
Wrong. These tests only verify that the sensor is operational but doesn’t mean the unit is detecting gas. For example, the sensor might be covered by a chemical which is preventing it from detecting gas, but the sensor could still be working.
Myth 3: I barely use my gas detector, so I don’t need to calibrate or setup a bump test procedure
Incorrect. Everything needs to be documented. If you don’t have the proper documentation that shows tests were done, then it doesn’t matter. Regardless of how many times you use the unit, you need to document and conduct calibrations and tests. You should always maintain a bump test schedule and document all that information.
This concludes my 3 part series on gas detector safety. Let me know if you have any ideas for other safety series!
source: OH&S







